Installing The Exposure Switch - Planmeca proline xc Installation Manual

Hide thumbs Also See for proline xc:
Table of Contents

Advertisement

3.2 Installing the exposure switch

Exposure switch holder
Body
Female
plug
Exposure
switch
cable
Exposure
switch
cable
Male
plug
Installation manual
Cover
Bottom of
column
The exposure switch must be located at least two meters
(7 feet) from the X-ray in a position that satisfies the local
radiation safety regulations regarding to operator safety
and the shielding of X-ray equipment.
Remove the cover from the exposure switch holder, note
that there is a ground lead connected to the cover, and
use the body as a template to mark and drill the two
securing holes. Secure the holder in position and replace
the cover.
Connect the female plug at one end of the exposure
switch cable to the socket on the underside of the expo-
sure switch holder.
Connect the male plug at the other end of the exposure
switch cable to one of the sockets at the bottom of the X-
ray column. It does not matter which socket the cable is
connected to.
Planmeca Proline XC X-ray unit 11
INSTALLING THE X-RAY

Advertisement

Table of Contents
loading

Table of Contents