Planmeca proline xc Installation Manual page 31

Hide thumbs Also See for proline xc:
Table of Contents

Advertisement

Cassette
holder
Beam
alignment
tool
Installation manual
Radiation beam
4
5
PLANMECA PM 2002 Proline Cephalostat 18x24
No. 652232
NOTE The soft tissue filter will filter a narrow area of the
4
5
PLANMECA PM 2002 Proline Cephalostat 18x24
No. 652232
Select exposure values of approximately 62kV, a milliam-
pere value of 6 and an exposure time of about one sec-
ond. The actual values will depend on how dark the room
is.
Stand behind the tube head and protect yourself from
radiation. Press and hold down the exposure button. The
image of the radiation beam from collimator number 4 will
appear on the alignment tool.
The radiation beam must appear within the borders of the
rectangle marked on the alignment tool. If it does not
refer to the Planmeca Proline XC Pan/Ceph X-ray unit
technical manual for information on how to align it.
left side of the radiation beam.
Slide the beam alignment tool to the right of the cassette
holder.
Planmeca Proline XC X-ray unit 27
CHECKING THE ALIGNMENT
collimator 4