CHECKING THE ALIGNMENT
18 Planmeca Proline XC X-ray unit
Enter the test mode (See section "Test mode" on page
14) to switch the radiation off.
Press and hold down the exposure button to drive the
rotating unit to a position where you can safely view the
secondary collimator when standing behind the tube
head. Stop pressing the exposure button when the rotat-
ing unit is in a convenient position.
An error code will appear on the display. Clear the error
code from the display by touching the OK field.
Exit the Test mode.
On the main display, touch the kV / mA field.
The Select kV/mA display appears. Select a kilovolt and a
milliampere value high enough to enable the radiation beam
to be seen in a darkened room, for example 62kV and 6mA.
The actual values will depend on how dark the room is.
Installation manual