CHECKING THE ALIGNMENT
Enter the service mode (See section "Service mode" on
page 15). By touching the arrow down field below the list
you can see more features. From the list of Service set-
tings select Primary collimator.
The display shown below appears.
Darken the room sufficiently so that you will be able to
see the image of the radiation beam on the alignment tool
(it is fluorescent and will glow when the radiation beam
strikes it), but not so dark that you cannot see the borders
of the alignment rectangle.
Stand behind the tube head and protect yourself from
radiation. Press and hold down the exposure button.
The image of the radiation beam will appear on the align-
ment tool.
Installation manual
Planmeca Proline XC X-ray unit 19