INSTALLING THE EXPOSURE SWITCH

7
INSTALLING THE EXPOSURE SWITCH
NOTE The exposure switch must be located in a position that satisfies the local radiation
safety regulations with regards to operator safety and the shielding of X-ray
equipment. The exposure switch can be mounted on the wall, or it can be hung from
the hook located on the back of the moving column if a protected area is within reach.
NOTE The wiring options for a remote installation (European and US versions) are shown
in section 10 "DIAGRAMS" on page 27.
CAUTION Only the Planmeca cross connection spiral cable (Planmeca part no.
10001193) can be used between the ProOne X-ray unit and the exposure
switch. This cable can be lengthened with an RJ12 Coupler and a Modular
extension cable (6wire with 2xRJ12).
7.1

Movable exposure switch

Use the holder plate to mark the position of the exposure switch.
If the wall is made of concrete or brick, use the 4x30 screws and the 6x30 expansion
anchors. Drill ø6mm (0.23 in.), 20 mm (0.8 in.) in depth, securing holes and place the
expansion anchors onto the holes.
If the wall is made of wood, use the 4x30 screws. Do not use the expansion anchors with
wooden wall. Drill ø3 mm (0.11 in.), 20 mm (0.8 in.) in depth, holes for the attachment
screws.
Attach the holder plate to the wall with the two attachment screws. The exposure switch
can then be positioned on the holder plate.
Holder plate
18
Planmeca ProOne
Printed copies of this document are considered uncontrolled.
7535.13.Rev001
08.07.2018
A
Holder plate
B
Installation Manual