CHECKING THE ALIGNMENT
30 Planmeca Proline XC X-ray unit
Radiation beam
6
Positioning light 25mm
Positioning light 0mm
Soft tissue filter 0mm
NOTE The soft tissue filter will filter a narrow area of the
The radiation beam must appear within the borders of the
rectangle marked on the alignment tool. If it does not
refer to the Planmeca Proline XC Pan/Ceph X-ray unit
technical manual for information on how to align it.
left side of the radiation beam.
When all the collimators have been checked, and if nec-
essary adjusted, exit the service mode, see section "Ser-
vice mode" on page 15.
You must now check the position of the soft tissue filter.
Turn to section 4.4 "Planmeca Proline XC pan/ceph X-ray
- checking the position of the soft tissue filter" on page
31.
collimator 6
Installation manual