CHECKING THE ALIGNMENT
20 Planmeca Proline XC X-ray unit
The radiation beam must appear within the borders of the
rectangle marked on the alignment tool. If it does not
refer to the Planmeca Proline XC panoramic X-ray tech-
nical manual for information on how to align it.
UP
Radiation beam
DOWN
When the collimator has been checked, and if necessary
adjusted, exit the service mode, see section "Service
mode" on page 15.
You must now take an exposure to check that the patient
positioning mechanism is correctly aligned, turn to sec-
tion 4.6 "Checking the alignment of the patient positioning
mechanism" on page 34.