7.1.7 Measured by JP (10mL- (in container))

66
This is the test procedure for purified water,
sterile purified water or water for injection
contained in a container of 10 mL or more.
The value when the measured temperature is at
25C ±1C and the conductivity change for 5
minutes is 0.1 S/cm or less is judged whether it
exceeds the evaluation standard, 5.0 S/cm or
not.
1.
Select the JP in the COND pharmacopeia mode
screen.
2.
Tap
on the right of the 10mL- item.
The changed settings are applied.
3.
Tap
to proceed to the next procedure.
4.
Immerse the COND electrode in sample solution
and tap
When measurement is completed, the conduc-
tivity of the sample solution and the measure-
ment condition are displayed as a measurement
result.
Tap
to return to the COND pharmaco-
peia mode screen.
to start the measurement.