4.3 SAL Measurement Setting

This section describes the procedures to set the conditions of SAL measurement.
Salinity concentration is calculated (Practical Salinity Scale (UNESCO 1978)) from the
measured value of conductivity.
Therefore, when the cell constant is set in conductivity measurement, there is no need to
input the cell constant. If no cell constant is set, refer to "3.1.1 Automatic Calibration
Setting" (P.35).

4.4 SAL Measurement Unit Setting

4.5 Temperature Setting

The settings of temperature compensation and temperature conversion in COND
measurement apply for SAL measurement (refer to "3.2.3 Temperature Setting" (P.41) and
"3.2.4 Temperature Conversion Function Setting" (P.42)).
50
1.
Tap the channel setting and the measurement
item in the MEAS screen to set "CH2" and "SAL".
2.
Tap
and tap "CH2 MEAS SET".
The SAL measurement setting items are
displayed.
3.
Select items and set the conditions.
The setting procedures for each item are
explained below.
The measurement unit of SAL measurement
values select either PPT or %.
1.
Tap
on the right of the Unit item.
2.
Select PPT or %.
The selected unit applies.