Chapter 7 Application Mode; Pharmacopeia Mode - Horiba Scientific DS-72G Instruction Manual

Cond meter
Table of Contents

Advertisement

Chapter 7 Application Mode

The application mode enables the measurement for the pharmaceutical water inspection
methods under various Pharmacopeias by conductivity measurement in conformance to
specific measurement methods. By simply submerging the electrode to a sample, the
instrument will walk you through the process and will determine the result. This chapter
explains about the settings and procedures of measurement using the pharmaceutical
water inspection methods under various Pharmacopeias by conductivity measurement.

7.1 Pharmacopeia Mode

In this mode, evaluation of pharmaceutical water (purified water and injection syringe
water) in conformity with US Pharmacopeia (USP), European Pharmacopeia (EP),
Japanese Pharmacopeia (JP), and Pharmacopoeia of the People's Republic of China
(PPRC) can be evaluated. This mode enables evaluation of pharmaceutical water that is
measured based on the standards in accordance with the Pharmacopeia regulations in
each country.
This mode has the function to indicate "Exceeded Limit" which shows that the sample
does not conform to the specifications when a measurement value does not satisfy the
Pharmacopeia regulations during measurement. When a measurement value is out of the
specification after the measurement, the non-conformity is indicated in the measurement
results. This applies for printouts.
This mode, you can save the measurement results only into a USB memory and print out
them. If you need to save or print out the data, turn ON the "Simultaneously Memory" of
"2.6 USB Memory Setting" (P.20) or "Auto Printout" of "2.7 Printer Setting" (P.22) in
advance.