3.2.3 Temperature Setting

If the temperature terminals of the instruction and electrode are not connected,
temperature setting is performed in MTC even when ATC is set.
Solution Temperature Entry
in MTC (Manual Temperature Compensation)
There are two types of temperature setting for
COND measurement; Automatic Temperature
Compensation (ATC) and Manual Temperature
Compensation (MTC).
In ATC, the instrument detects the solution
temperature with the connected temperature
sensor, and performs temperature
compensation for the COND values of the
standard solutions used for calibration.
In MTC, measure the solution temperature and
enter the value in advance. The instrument
performs temperature compensation using the
entered temperature.
1.
Display the Temperature item when select MTC.
Tap
on the right of the Temperature item.
2.
Enter the solution temperature on the numerical-
key screen.
3.
Tap
.
The setting applies.
To cancel the settings, tap
.
41