Download Print this page
Horiba Scientific LAQUA DO220 Quick Manual

Horiba Scientific LAQUA DO220 Quick Manual

Handheld meters

Advertisement

Quick Links

Quick Guide
LAQUA 200 Series
DO210/220 & PD210/220
Handheld Meters
DO210
www.horiba-laqua.com
4. Press ENT button to confirm the DO reading. To abort calibration, press
MEAS button.
The meter will switch to measurement
mode and the electrode icon will
appear on the screen.
5. Immerse the DO electrode in zero solution (5% sodium sulfite) and repeat step
nos. 3 - 4 to perform zero calibration (optional).
Make sure that the membrane and temperature sensor of the DO electrode are fully immersed in the
zero solution.
In step no. 3, CAL button must be pressed twice so that the meter will switch to zero calibration mode.
6. Rinse the DO electrode with clean water and blot dry with lint-free tissue.
7. Immerse the DO electrode in sample.
Make sure that the membrane and
temperature sensor of the DO electrode are
fully immersed in the sample.
Dissolved Oxygen Calibration & Measurement
1. Rinse the DO electrode
with clean water and blot
dry with lint-free tissue.
There should be no water droplets on
the membrane and temperature sensor.
2. Hold the DO electrode still in clean air.
Perform air calibration in clean air at
a location not subjected to dramatic
temperature change, rain, or direct wind.
Two calibration modes are available in the
DO meter:
1. Saturated oxygen concentration mode (%)
2. Dissolved oxygen measurement mode
(mg/L)
The default mode is the former. Press MODE
button to switch to the latter.
3. Press CAL button on the meter to switch
to calibration mode.
The  will start blinking until the reading stabilizes.
Press
Press

Advertisement

loading

Summary of Contents for Horiba Scientific LAQUA DO220

  • Page 1 Dissolved Oxygen Calibration & Measurement 1. Rinse the DO electrode with clean water and blot Quick Guide dry with lint-free tissue. LAQUA 200 Series There should be no water droplets on DO210/220 & PD210/220 the membrane and temperature sensor. Handheld Meters 2.
  • Page 2 Dissolved Oxygen Dissolved Oxygen Measurement Modes Calibration Calibration Data Auto Stable Auto Hold Real Time VIDEO VIDEO VIDEO VIDEO VIDEO Dissolved Oxygen Setup P4 CLK* • P4.2 TIME - Set time • P4.1 DATE - Set date P3 GEN • P3.4 RSET - Reset meter •...

This manual is also suitable for:

Laqua do210Laqua pd210Laqua 200 seriesLaqua pd220