8.1.2 COND Checker (X-52) Mode

In this mode, the instrument operations are checked by using the optional checker X-52.
Refer also to the instruction manual of the checker X-52 before the operation.
When the COND periodical check mode (X-52) starts, the following items are set
automatically as follows.
・ Unit: S/m
・ Cell constant: 1.000  100 m
・ Temperature setting: ATC
Follow the guidance to check.
Span check
19.00 S/m
1.900 S/m
190.0 mS/m
19.00 mS/m
1.900 mS/m
Linearity check
10.00 S/m
1.000 S/m
100.0 mS/m
10.00 mS/m
1.000 mS/m
0.000 mS/m
Temperature check
0.0C
30.0C
60.0C
100.0C
The conductivity measurement values displayed during in the above operations are not
concerned with measurement.
When all the check is completed, the result is displayed automatically.