7.1.6 Measured by JP (0mL-10mL (in container))

This is the test procedure for purified water,
sterile purified water or water for injection
contained in a container of 10 mL or less.
The value when the measured temperature is at
25C ±1C and the conductivity change for 5
minutes is 0.1 S/cm or less is judged whether it
exceeds the evaluation standard, 25 S/cm or
not.
1.
Select the JP in the COND pharmacopeia mode
screen.
2.
Tap
on the right of the 0mL-10mL item.
The changed settings are applied.
3.
Tap
to proceed to the next procedure.
4.
Immerse the COND electrode in sample solution
and tap
When measurement is completed, the conduc-
tivity of the sample solution and the measure-
ment condition are displayed as a measurement
result.
Tap
to return to the COND pharmaco-
peia mode screen.
to start the measurement.