Keysight Technologies N8262A P Series User Manual page 136

Modular power meter
Table of Contents

Advertisement

4
Maintenance
–321
136
System error; Dty Cyc may impair accuracy with ECP sensor
310
This indicates that the sensor connected is for use with CW signals only.
System error; Ch A Dty Cyc may impair accuracy with ECP sensor
310
This indicates that the sensor connected to Channel A is for use with CW signals only.
System error; Ch B Dty Cyc may impair accuracy with ECP sensor
310
This indicates that the sensor connected to Channel B is for use with CW signals only.
System error; Sensor EEPROM Read Failed - critical data not found or unreadable
310
This indicates a failure with your E-Series Power Sensor. Refer to your power sensor
manual for details on returning it for repair.
System error; Sensor EEPROM Read Completed OK but optional data block(s) not
found or unreadable
310
This indicates a failure with your E-Series Power Sensor. Refer to your power sensor
manual for details on returning it for repair.
System error; Sensor EEPROM Read Failed - unknown EEPROM table format
310
This indicates a failure with your E-Series Power Sensor. Refer to your power sensor
manual for details on returning it for repair.
System error; Sensor EEPROM < > data not found or unreadable
Where < > refers to the sensor data block covered, for example, Linearity, Temp - Comp
310
(temperature compensation).
This indicates a failure with your E-Series Power Sensor. Refer to your power sensor
manual for details on returning it for repair.
Out of memory
The power meter required more memory than was available to run an internal
operation.
Self-test Failed;
The -330, "Self-test Failed" errors indicate that you have a problem with your power
330
meter. Refer to
do with your faulty power meter.
330
Self-test Failed; Measurement Channel Fault
330
Self-test Failed; Measurement Channel A Fault
Self-test Failed; Measurement Channel B Fault
330
Self-test Failed; Calibrator Fault
330
Refer to
"Calibrator"
"Contacting Keysight Technologies"
on page 128 if you require a description of the calibrator test.
on page 139 for details of what to
Keysight N8262A User's Guide