Keysight Technologies N8262A P Series User Manual page 134

Modular power meter
Table of Contents

Advertisement

4
Maintenance
–230
134
Illegal parameter value
A discrete parameter was received which was not a valid choice for the command. You
224
may have used an invalid parameter choice.
For example, TRIG:SOUR EXT.
Lists not same length
226
This occurs when SENSe:CORRection:CSET[1]|CSET2:STATe is set to ON and the
frequency and calibration/offset lists do not correspond in length.
Data corrupt or stale
This occurs when a FETC? is attempted and either a reset has been received or the
230
power meter state has changed such that the current measurement is invalidated (for
example, a change of frequency setting or triggering conditions).
Data corrupt or stale; Please zero and calibrate Channel A
When CAL[1|2]:RCAL is set to ON and the sensor currently connected to Channel A has
230
not been zeroed and calibrated, then any command which would normally return a
measurement result (for example FETC?, READ?, or MEAS?) will generate this error
message.
Data corrupt or stale; Please zero and calibrate Channel B
When CAL[1|2]:RCAL is set to ON and the sensor currently connected to Channel B has
230
not been zeroed and calibrated, then any command which would normally return a
measurement result (for example FETC?, READ?, or MEAS?) will generate this error
message.
Data corrupt or stale; Please zero Channel A
When CAL[1|2]:RCAL is set to ON and the sensor currently connected to Channel A has
230
not been zeroed, then any command which would normally return a measurement
result (for example FETC?, READ?, or MEAS?) will generate this error message.
Data corrupt or stale; Please zero Channel B
When CAL[1|2]:RCAL is set to ON and the sensor currently connected to Channel B has
not been zeroed, then any command which would normally return a measurement
result (for example FETC?, READ?, or MEAS?) will generate this error message.
Data corrupt or stale; Please calibrate Channel A
When CAL[1|2]:RCAL is set to ON and the sensor currently connected to Channel B has
230
not been calibrated, then any command which would normally return a measurement
result (for example FETC?, READ?, or MEAS?) will generate this error message.
Data corrupt or stale; Please calibrate Channel B
When CAL[1|2]:RCAL is set to ON and the sensor currently connected to Channel B has
230
not been calibrated, then any command which would normally return a measurement
result (for example FETC?, READ?, or MEAS?) will generate this error message.
Keysight N8262A User's Guide