Keysight Technologies N8262A P Series User Manual page 132

Modular power meter
Table of Contents

Advertisement

4
Maintenance
–151
132
Invalid character in number
121
An invalid character was found in the number specified for a parameter value.
For example, SENS:AVER:COUN 128#H.
Exponent too large
123
A numeric parameter was found whose exponent was larger than 32,000.
For example, SENS:COUN 1E34000.
Too many digits
124
A numeric parameter was found whose mantissa contained more than 255 digits,
excluding leading zeros.
Numeric data not allowed
A numeric value was received within a command which does not accept a numeric
128
value.
For example, MEM:CLE 24.
Invalid suffix
A suffix was incorrectly specified for a numeric parameter. You may have misspelled the
131
suffix.
For example, SENS:FREQ 200KZ.
Suffix too long
134
A suffix used contained more than 12 characters.
For example, SENS:FREQ 2MHZZZZZZZZZZZ.
Suffix not allowed
138
A suffix was received following a numeric parameter which does not accept a suffix.
For example, INIT:CONT 0Hz.
Character data not allowed
A discrete parameter was received but a character string or a numeric parameter was
148
expected. Check the list of parameters to verify that you have used a valid parameter
type.
For example, MEM:CLE CUSTOM_1.
Invalid string data
An invalid string was received. Check to see if you have enclosed the character string in
single or double quotes.
For example, MEM:CLE "CUSTOM_1.
String data not allowed
A character string was received but is not allowed for the command. Check the list of
158
parameters to verify that you have used a valid parameter type.
For example, LIM:STAT 'ON'.
Keysight N8262A User's Guide