Switch Valves (Function 85) - Siemens FIDAMAT 6-G Operating Instructions Manual

Gas analyzer for measuring total hydrocarbon concentrations
Table of Contents

Advertisement

Press softkey 1 again to carry out the interference gas correction via the serial interface RS
485 (ELAN).
Here, the same inputs are to be made as for the interference gas correction
input"
Also input the channel number and component number of the interference gas analyzer. In
the display, the type of gas allocated to the channel and component as well as the device
status appear.
With softkey 2, you can set whether the interference gas correction should only apply for a
certain measuring range.
See also
ELAN configuration (function 73) (Page 5-49)
5.11.16

Switch valves (function 85)

85 Switch valves
Figure 5-65
With this function, you can manually switch up to six valves. This is done via the relays which
are allocated to the individual valves, which are available on the motherboard and optional
board. The precondition is that the corresponding relay was previously configured with
function 71 ("relay allocation")
The "Switch valves" function only applies for the relay functions "zero gas", "test gas 1 to 4"
and "sample gas".
Only one valve can be switched at a time, since the valves are interlocked against each
other.
See also
Relay allocation (function 71) (Page 5-40)
FIDAMAT 6-G
Operating instructions, 07/2006, A5E00478463
.
1
Sample pt.6
Rel. 1
2
Sample pt.2
Rel. 2
3
Sample pt.4
Rel. 3
Switching valves
THC
Operation
5.11 Configuration
"...via analog
5-59