Installation Requirements For The Fidamat 6-G - Siemens FIDAMAT 6-G Operating Instructions Manual

Gas analyzer for measuring total hydrocarbon concentrations
Table of Contents

Advertisement

Notice
After performing servicing work on the containment system, always carry out a leak check!
Pressure sensors
The FIDAMAT 6 has two internal pressure sensors for measuring the sample gas pressure
(= combustion air pressure) and the combustion gas pressure.
See also
Gas flow of the FIDAMAT 6-G (Page 2-8)
Gas and electric connections (Page 3-6)
Checking for leaks (Page 4-3)
Technical data of the FIDAMAT 6 series (Page 7-1)
Technical data of the FIDAMAT 6-G (Page 7-3)
3.2.2

Installation requirements for the FIDAMAT 6-G

Sample gas line
For the gas supply and discharge lines, there is a directly heated input point (sample) and a
directly heated output point for each. Separately heated sample gas supply lines require their
own power supply and temperature regulation. The sample gas is supplied unheated.
The hexagon of the bulkhead unions sits in a turn lock. When tightening the screw caps at
the screwed connections make sure that you properly plug with a suitable open-ended
spanner. Otherwise there is a danger that the gas path will leak.
FIDAMAT 6-G
Operating instructions, 07/2006, A5E00478463
Installation and connection
3.2 Installation requirements
3-5