Autocal (Function 24) - Siemens FIDAMAT 6-G Operating Instructions Manual

Gas analyzer for measuring total hydrocarbon concentrations
Table of Contents

Advertisement

Operation
5.8 Calibration
5.8.6

Autocal (function 24)

Note
You can only make use of automatic calibration (Autocal) if your device contains additional
electronics (optional board). If this isn't the case, a corresponding message will appear on
the display after the Autocal function is called.
Automatic calibration can only be started if the device is in the "measure" operating state!
Figure 5-13
Note
The settings for "Autocal/Check..." (subfunctions 1 to 3) are valid both for Autocal (automatic
calibration) as well as for Autocal Check (automatic check that the set calibration tolerances
are complied with, without calibration).
The settings for "Autocal Check" (4th subfunction) are only with reference to the check of the
calibration tolerances without calibration.
5-16
Autocal function
Operating instructions, 07/2006, A5E00478463
FIDAMAT 6-G