Also enter the channel number and component number of the interfering gas analyzer. The
type of gas assigned to the channel and component as well as the device status appear in
the display.
Note
On the device supplying the correction data, the parameter
(function 73) must be set to
With softkey 2, you set whether the cross-interference correction is to apply only to certain
measuring ranges.
See also
ELAN configuration (function 73) (Page 115)
8.6.15

Switch valves (function 85)

Figure 8-61
With this function, you can manually switch up to six valves. This is done via the relays which
are allocated to the individual valves and are available on the motherboard and optional
board. The precondition is that the corresponding relays have been previously configured
with function 71 ("Relay outputs")
The "Switch valves" function only applies to the relay functions "Zero gas", "Calibration gas 1
to 4" and "Sample gas".
Only one valve can be switched at a time, since the valves are mutually interlocked.
See also
Relay outputs (function 71) (Page 107)
FIDAMAT 6
Operating Instructions, 01/2019, A5E00222135-04
"On"
.
Switching valves
8.6 Configuration
"Measured value telegram"
Functions
125