Installing Fittings and Capillaries

Installing Fittings and Capillaries

Solvent can spray under high pressure.
WA R N I N G
Deformation of fittings and seals
C A U T I O N
Liquid drops under high pressure up to 1200 bar act like solid parts. Tightening
connections under high pressure can deform or destroy fittings and seals.
The lifetime of a fitting depends on how firmly it has been tightened; firm tightening
N O T E
reduces the lifetime.
If fitting has been overtightened, replace it.
1 Install fittings and capillaries.
2 Tighten fittings and capillaries.
130
Observe appropriate safety procedures (for example, goggles, safety gloves and
protective clothing), when opening flow path.
Never tighten flow connections under pressure.
1290 Infinity Quaternary Pump User Manual