2-1. Preparing for Measurement
Setting the Upper Limit of Pressurizing
Specify the upper limit of pressurizing settings for the right ankle and left ankle.
Normally "Auto" is selected. When "Auto" is set, the system pressurizes the cuff and measures the
patient's blood pressure automatically. If the patient complains of discomfort due to cuff
pressurizing, change the setting to "Manual" and set an upper limit of pressurizing.
1.
Press the [MAX PRESSURE] button.
2.
Select "Auto" or "Manual" for the right leg.
If you selected [Auto], go to step 4.
3.
Set the upper limit of pressurizing.
The input range is 100 mmHg to 280 mmHg.
4.
Repeat steps 2 and 3 for the left leg.