5-9. Guidance and Manufacturer's Declaration

Warning
The BP-203RPE3 is intended for use in the electromagnetic environment specified below. The
customer or the user of the BP-203RPE3 should assure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Use only the specified supplies for the cord, cuff, and other parts. Do not install other than
specified options. Otherwise and accident may result.
Electromagnetic emissions IEC60601-1-2
Compliance
Group 1
Class A
Class A
Complies
Electromagnetic immunity IEC60601-1-2
IEC 60601
test level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV differential
mode
±2 kV common mode
<5 % U
T
(>95 % dip in U
)
T
for 0.5 cycle
70 % U
T
(30 % dip in U
)
T
for 25 cycle
<5 % U
T
(>95 % dip in U
)
T
for 5 sec
3 A/m
Electromagnetic environment – guidance
The BP-203RPE3 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
The BP-203RPE3 is suitable for use in all establishments other
than domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used for
domestic purposes.
Compliance level
±6 kV contact
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
±8 kV air
material, the relative humidity should be at
least 30 %.
±2 kV for power
Mains power quality should be that of a
supply lines
typical commercial or hospital environment.
±1 kV for input/output
lines
±1 kV differential
Mains power quality should be that of a
mode
typical commercial or hospital environment.
±2 kV common mode
<5 % U
Mains power quality should be that of a
T
typical commercial or hospital environment.
(>95 % dip in U
)
T
If the user of the BP-203RPE3 requires
for 0.5 cycle
continued operation during power mains
interruptions, it is recommended that the
70 % U
T
BP-203RPE3 be powered from an
(30 % dip in U
)
uninterruptible power supply or a battery.
T
for 25 cycle
<5 % U
T
(>95 % dip in U
)
T
for 5 sec
3 A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital
environment.
Electromagnetic environment -
guidance