Omron Colin VP-1000 plus Manual page 143

Non-invasive vascular screening device
Table of Contents

Advertisement

Portable and mobile RF communications equipment and the BP-203RPE 3
The
is intended for use in an electromagnetic environment in which radiated RF
BP-203RPE3
disturbances are controlled. The customer or the user of the
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the
maximum output power of the communications equipment.
Rated Maximum Output Power of
Transmitter in watt
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note: At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
Cables and Sensors
AC power cord
AC power cord
LAN cable
USB cable
Arm Cuff Hose
Ankle Cuff Hose
ECG lead cable
PCG cable
Sensor cable
Table-tap
Recommended separation distance between
Separation distance according to frequency of transmitter in meter
150 kHz to MHz
d
=
1.2 p
0.12
0.38
1.2
3.8
12
Cables IEC 60601-1-2
Maximum
Length
0.8 m
- RF emissions, CISPR 11, ClassA / Group 1
1.1 m
- Harmonic emissions, IEC61000-3-2
10.0 m
- Voltage fluctuations/flicker emission, IEC61000-3-3
1.2 m
- Electrostatic discharge (ESD), IEC61000-4-2
2.4 m
- Electric fast transient/burst, IEC61000-4-4
0.5 m
- Surge, IEC61000-4-5
2.3 m
- Conducted RF, IEC61000-4-6
2.0 m
- Radiated RF, IEC61000-4-3
3.0 m
3.5 m
5-9. Guidance and Manufacturer's Declaration
BP-203RPE3
as recommended below, according to the
BP-203RPE3
80 MHz to 800 MHz
d
=
1.2 p
0.12
0.38
1.2
3.8
12
Complies with
can help prevent
800 MHz to 2.5GHz
d
=
2.3 p
0.23
0.73
2.3
7.3
23
141

Advertisement

Table of Contents
loading

This manual is also suitable for:

Colin bp-203rpe iii

Table of Contents