5-9. Guidance and Manufacturer's Declaration
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the BP-203RPE3 is used exceeds the
applicable RF compliance level above, the BP-203RPE3 should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the
BP-203RPE3.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
140
IEC 60601 test level
3Vrms
3Vrms
150 kHz to 80 MHz
80%AM (2Hz)
3 V/m
3 V/m
80 MHz to 2.5 GHz
80%AM (2Hz)
Compliance level
Portable and mobile RF communications
equipment should be used no closer to any
part of the BP-203RPE3, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
d
d
d
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Electromagnetic environment -
guidance
1.2 P
=
1.2 P
80 MHz to 800 MHz
=
2.3 P
800 MHz to 2.5 GHz
=
d
is the recommended separation
a
should be less than the compliance