Summary of Contents for Omron MicroAIR U100

  • Page 1 Instruction manual MESH NEBULIZER MicroAIR U100 (NE-U100-E) • English • Français • Deutsch • Italiano • Español • Nederlands • Русский ‫• اﻟﻌرﺑﯾﺔ‬ IM-NE-U100-E-04-08/2017 3125434-0D...
  • Page 2: Table Of Contents

    How to assemble the unit and fill the medication container. . . 14 How to inhale with the OMRON MicroAIR U100 ... 17 How to clean the unit after each inhalation ....19 How to disinfect the parts .
  • Page 3: Intended Use

    • The user should also be capable of understanding general operation of MicroAIR U100 and the content of instruction manual. Intended Patients This product should not be used by patients, who are unconscious or are not breathing spontaneously.
  • Page 4: Important Safety Precautions

    Important Safety Precautions This section provides information about how to use the product safely and to prevent injury to yourself and others as well as property damage. Indicates a potentially hazardous situation which, if not avoided, could result in Warning: death or serious injury.
  • Page 5 • Do not operate the nebulizer with the medication container empty. • Approved for human use only. • Use only OMRON original and optional accessories specified for this device. • Do not add more than 10 ml of medication to the medication container.
  • Page 6 Important Safety Precautions (Power source) • Use only the original AC adapter. • Fully insert the power plug into the outlet. • Wipe the dust off from the power plug. • When handling the AC adapter, take care not to do the following: Do not damage.
  • Page 7: Features And Benefits

    Features and benefits In the MicroAIR U100, aerosol is generated by pushing liquid medication through fine holes in the metal alloy mesh. This is realised by a titanium vibrating element, oscillating at high frequency. True portable, pocket size The device will fit in any handbag or briefcase. Running up to approx. 4 hours on just two AA/LR6 batteries, your medication is always within reach wherever you go, whatever you do.
  • Page 8: Contents Of The Omron Microair U100 Package

    Contents of the OMRON MicroAIR U100 package The following items are contained in the package: Medication Main Unit Cover Main Unit Container Mesh Cap Mask Adapter Mouthpiece Soft Pouch Child Mask (PVC) Adult Mask (PVC) Batteries (2x AA/LR6) • Instruction Manual...
  • Page 9: Names And Functions Of The Parts

    Names and functions of the parts Mesh Cap Contains metal alloy mesh for nebulization. Mesh Nebulized medication exits here through minute pores. Medication container Air opening Stabilizes nebulization. Medication container lid release button Medication container can be opened when the buttons on both sides are pressed.
  • Page 10 Names and functions of the parts Rear View Battery Cover release button Battery Cover Main unit cover Protects the Main Unit with the attached Medication Container and Mesh Cap during storage.
  • Page 11: Before Using The Product For The First Time

    Before using the product for the first time Before first use: I ) Carefully read the Instruction Manual II ) Disassemble, clean and disinfect the parts Disassembly...
  • Page 12: How To Insert Or Replace The Batteries

    How to insert or replace the batteries Inserting batteries 1. Remove the battery cover on bottom side by pushing on the release buttons on both sides. 2. Insert the batteries according to the battery polarity indicated inside the battery compartment. 3.
  • Page 13: Using The Ac Adapter (Optional)

    Using the AC adapter (Optional) Caution: Use only the original AC adapter as indicated in section “Optional Medical Accessories”. To connect the AC adapter to the main unit (1) Remove the battery cover and remove the batteries, see page 12. (2) Place the main unit on the AC adapter connection stand as shown in the illustration on the right.
  • Page 14: How To Assemble The Unit And Fill The Medication Container

    How to assemble the unit and fill the medication container Clean and disinfect the product’s parts before using the product for the first Caution: time, after the product has not been used for an extended period of time and when parts are contaminated. (Refer to page 19 for instructions.) 1.
  • Page 15 How to assemble the unit and fill the medication container 4. Fill the medication container as illustrated. The maximum capacity is 10 ml. Mesh 5. Place the mesh cap. Note: Do not touch the mesh. 6. Push the lid of medication container into place until it clicks. Mesh Cap is fixed.
  • Page 16 How to assemble the unit and fill the medication container 7. Install the mask adapter on the main unit. 8. Install the mask or mouthpiece on the mask adapter. The device is now ready for use. Refer to the next section how to inhale.
  • Page 17: How To Inhale With The Omron Microair U100

    How to inhale with the OMRON MicroAIR U100 Caution: For type, dose, and regime of medication, follow the Medication instructions of your doctor or pharmacist. 1. Slightly tilt the unit as illustrated in the figure. In this position, the vibrating element is immersed in the medication and nebulization will start after power on.
  • Page 18 How to inhale with the OMRON MicroAIR U100 Note: If high-viscosity medication is used, nebulization may be reduced. If excessive medication collects on the mesh, nebulization may stop. In that case, turn off the power and remove the mask adapter. Absorb the medication by means of a lint-free cloth.
  • Page 19: How To Clean The Unit After Each Inhalation

    How to clean the unit after each inhalation The following items should be disassembled and cleaned after each use: Medication container, mesh cap, mask adapter, mouthpiece and mask. Note: if the device is not cleaned and disinfected correctly and frequently, micro-organisms may remain in the unit, causing risk of infection.
  • Page 20 How to clean the unit after each inhalation 4. Fill the medication container with water and nebulize the water. To prevent medication from drying and adhering to the mesh Water after use, nebulize the water for 1 to 2 minutes. Do not inhale water.
  • Page 21 How to clean the unit after each inhalation 6. Follow the below instructions to clean all parts. • Mesh Cap Wash the mesh cap in mild (neutral) detergent, then rinse off with clean water. Remove excess water and allow to air dry in a clean place.
  • Page 22: How To Disinfect The Parts

    How to disinfect the parts Note: if the device is not cleaned and disinfected correctly and as frequently as indicated, microorganisms may remain in the unit, causing risk of infection. The Medication Container, Mesh Cap, Mouthpiece, Mask and Mask Adapter should be disinfected before using the unit for the first time, after the unit has not been used for a long period of time, or daily during normal use.
  • Page 23 How to disinfect the parts Disinfection by commercially available disinfectant Use a commercially available disinfectant. Follow the instructions provided by the disinfectant manufacturer. Note: Never clean with benzene or thinner. • Vibrating element and its surrounding Soak the vibrating element and the area around the vibrating element in disinfectant.
  • Page 24: How To Replace The Mesh Cap

    How to replace the mesh cap The mesh cap is a consumable part. OMRON recommends to exchange the mesh cap after approximately 1 year (Please refer to page 3 “Durable period”). Remove the mesh cap and install a new one as described on page 14 (section “How to assemble the unit and fill the medication container”).
  • Page 25: Troubleshooting

    Troubleshooting In case of any of the below problems occur during use, first check that no other electrical device is within 30cm. If the problem persists, please refer to the table below. Trouble Possible cause Corrective action Batteries are low (the orange Replace the batteries with new battery low indicator flashes).
  • Page 26 Treatment times may vary among indicator turns on but the medication used. medications and patients. it nebulizes weakly, or is taking too long for treatment. If the unit does not nebulize properly after checking the possible causes above, contact your OMRON dealer.
  • Page 27: Technical Data

    Technical data Product Category: Nebulizers Product Description: Mesh Nebulizer Model (code): MicroAIR U100 (NE-U100-E) Rating: 100-240V 50/60Hz (Optional AC adapter) 3 VDC (two batteries/Type AA/LR6) Power Consumption: Approx. 1.2 W Residual Volume: Approx. 0.5 ml Battery Life: Approx. 4 hours...
  • Page 28 Notes: • Subject to technical modification without prior notice. • This OMRON product is produced under the strict quality system of OMRON HEALTHCARE Co., Ltd., Japan. • The device may not work if the temperature and voltage conditions are different to those defined in the specifications.
  • Page 29 Technical data Particle Size: *MMAD 4.5 μm MMAD = Mass Median Aerodynamic Diameter Medication Container Capacity: 10 ml maximum Sound: Approx. 20 dB Nebulization rate: **Greater than 0.25 ml/min (by weight loss) Aerosol Output: 0.5 ml (2 ml, 1%NaF) Aerosol Output Rate: 0.1 ml/min (2 ml, 1%NaF) Notes: •...
  • Page 30 This device manufactured by OMRON HEALTHCARE Co., Ltd. conforms to EN60601-1-2:2015 Electro Magnetic Compatibility (EMC) standard. Further documentation in accordance with this EMC standard is available at OMRON HEALTHCARE EUROPE at the address mentioned in this instruction manual or at www.omron-healthcare.com Correct Disposal of This Product (Waste Electrical &...
  • Page 31: Warranty

    This product is guaranteed by OMRON for a period of 3 years after the date of purchase. The proper construction, workmanship and materials of this product is guaranteed by OMRON. During this period of guarantee OMRON will, without charge for labour or parts, repair or replace the defect product or any defective parts.
  • Page 32 Manufacturer OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, KYOTO, 617-0002 JAPAN EU-representative OMRON HEALTHCARE EUROPE B.V. Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS www.omron-healthcare.com OMRON HEALTHCARE Co., Ltd. Matsusaka Factory Production facility 1855-370, Kubo-cho, Matsusaka-shi, Mie, 515-8503 Japan OMRON HEALTHCARE UK LTD.