HP 8924C Application Manual page 330

Cdma mobile station test set
Hide thumbs Also See for 8924C:
Table of Contents

Advertisement

Chapter 10, Controlling Program Flow
Controlling Program Flow Procedure
620
!
Data/Signal Register Group Enable Register must be set HIGH.
630
!
640
OUTPUT Inst_address;"STAT:QUES:ENAB 256"
650
!
660
! 3) Hardware Status Register #1 Group
670
!
Condition register conditions which will set the Summary Message
680
!
TRUE if they occur:
690
!
Bit 5: Measurement limits exceeded
700
!
Bit 4: Power-up Self-test failed
710
!
Bit 3: Overpower protection tripped
720
!
730
Hardware1_reg=32+16+8
740
!
750
!
Set the Transition Filters to allow only positive transitions in
760
!
the assigned condition(s) to pass to the Event Register
770
!
780
OUTPUT Inst_address;"STAT:HARD1:PTR";Hardware1_reg
790
OUTPUT Inst_address;"STAT:HARD1:NTR 0"
800
!
810
!
Set up the Hardware Status Register #1 Group Enable Register to
820
!
generate the Summary Message.
830
!
840
OUTPUT Inst_address;"STAT:HARD1:ENAB";Hardware1_reg
850
!
860
! 4) Set the correct Summary Message bit(s) in the Service Request
870
!
Enable Register to generate a Service Request (SRQ) if the
880
!
Summary Message(s) become TRUE.
890
!
Bit 5 = Standard Event Status Register Summary Message
900
!
910
!
Bit 3 = Questionable Data/Signal Register Group Summary Message
920
!
930
!
Bit 0 = Hardware Status Register #1 Group Summary Message
940
!
950
!
960
Srq_enab_reg=32+8+1
970
OUTPUT Inst_address;"*SRE";Srq_enab_reg
980
!
990
! 5) Set up the Active Controller to respond to an SRQ interrupt:
1000
!
Call subprogram Check_interrupt if an SRQ condition exists on select
1010
!
code 7. The interrupt priority level is set to 15 (highest level).
1020
!
1030
ON INTR 7,15 CALL Srvice_interupt
1040
!
1050
! 6) Enable interrupts on select code 7:
1060
!
The interface mask is set to a value of 2 which enables interrupts on
1070
!
the HP-IB bus when the SRQ line is asserted.
1080
!
1090
ENABLE INTR 7;2
1100
!
1110
! Start of the dummy loop:
1120
!
1130
LOOP
1140
DISP "I am sitting in a dummy loop."
1150
END LOOP
1160
!
1170
END
1180
!
1190 Srvice_interupt:SUB Srvice_interupt
1200
!
1210
OPTION BASE 1
1220
COM /Io_names/ INTEGER Inst_address,Std_event_reg,Calibration_reg
330
decimal value = 2^5 = 32
decimal value = 2^4 = 16
decimal value = 2^3 =
decimal value = 2^5 = 32
decimal value = 2^3 =
decimal value = 2^0 =
N:\mkt\MANUALS\HP8924C\APPMOD\BOOK\chapters\amsrq.fb