Table D-5. Additional Device Status (Command #48 - MSA Ultima X Series Installation & Maintenance Instructions Manual

Gas monitors
Hide thumbs Also See for Ultima X Series:
Table of Contents

Advertisement

Table D-5. Additional Device Status (Command #48)

Command #48 returns 5 bytes of data, with the following status information:
BYTE
BIT
MEANING
0
0
Configuration Reset
1
Main ram fault
2
Main flash fault
3
EEprom write error
4
Incompatible sensor
5
Sensor quick under range
6
Sensor UNDer range
7
Calibration fault
1
0
Sensor Missing
1
Sensor Overrange
2
Overrange Lock
3
Parameter Fault
4
Sensor Warm up
5
Sensor Config Reset
6
Sensor Power Fault
7
5V Power Fault
2
0
Zero Countdown
1
Apply Zero Gas
2
Span Countdown
3
Apply Span Gas
4
Cal Aborted
5
Zero Fault
6
Span Fault
7
Cal OK
D-4
CLASS
DEVICE
STATUS
BITS SET
Error
4,7
Error
4,7
Error
4,7
Error
4,7
Error
4,7
Error
4,7
Error
4,7
Error
4,7
Error
4,7
Warning
Warning
Error
4,7
Warning
Warning
Error
Error
Info
Info
Info
Info
Info
Info
Info
Info