Command #180: Write Sensor Swap Delay Enable

This command writes command number to the Ultima X Gas Monitor to
enable or disable the two-minute swap delay feature. This device
feature enables a two-minute hold-off of the sensor missing fault,
allowing the user to "Swap" or change sensor modules without having
the 4-20 mA set to the fault condition. The configuration change bit will
be set, and the configuration change counter will be incremented.
Request Data Bytes
BYTE
FORMAT
0
Enum
Response Data Bytes
BYTE
FORMAT
0
Enum
Command-Specific Response Codes
CODE
CLASS
0
Success
1-2
3
Error
4
5
Error
6
7
Error
8-15
16
Error
17-31
32
Error
33-127
DESCRIPTION
Swap Delay 1 = Enable 0 = Disable
DESCRIPTION
Swap Delay 1 = Enable 0 = Disable
DESCRIPTION
No Command-Specific Errors
Undefined
Parameter too large
Undefined
Too few data bytes
Undefined
In write protect mode
Undefined
Access Restricted
Undefined
Busy
Undefined