Command #177: Write Upper Trim Point - MSA Ultima X Series Installation & Maintenance Instructions Manual

Gas monitors
Hide thumbs Also See for Ultima X Series:
Table of Contents

Advertisement

Command #177: Write Upper Trim Point

Writes the Ultima X Upper Trim or Span point value. The Ultima Gas
Monitor uses the Upper trim point value to perform Span calibration.
When a Span calibration is performed, the device automatically sets the
highest reading obtained to this Span value. The adjustment range on
the Upper Trim Point is one display unit to the full-scale limit.
Request Data Bytes
BYTE
FORMAT
0-3
Float
Response Data Bytes
BYTE
FORMAT
0-3
Float
Command-Specific Response Codes
CODE
CLASS
0
Success
1-2
3
Error
4
Error
5
Error
6
7
Error
8-15
16
Error
17-31
32
Error
33-127
D-18
DESCRIPTION
Upper Trim Point (Span) Value
DESCRIPTION
Upper Trim Point (Span) Value
DESCRIPTION
No Command-Specific Errors
Undefined
Parameter too large
Parameter too small
Too few data bytes
Undefined
In write protect mode
Undefined
Access Restricted
Undefined
Busy
Undefined

Hide quick links: