Command #187: Write Alert Option - MSA Ultima X Series Installation & Maintenance Instructions Manual

Gas monitors
Hide thumbs Also See for Ultima X Series:
Table of Contents

Advertisement

Command #187: Write Alert Option

This command disables or enables the Alert Option on the Ultima X
unit. This is a single byte containing a 0 if disabled or 1 if enabled. If
enabled, the Alert Option will cause the 4-20 mA to be set to 3.75 mA
during calibration of an Oxygen sensor (if the Cal Signal Option is also
enabled). If the Alert Option is disabled and the Cal Signal enabled, the
output will be set to 21 mA during Oxygen sensor calibration.
Calibration
Power on RESET
(Countdown)
4 - 20 CAL mA
(Oxygen)
4 -20 POR mA (Oxygen)
Request Data Bytes
BYTE
FORMAT
0
Enum
Response Data Bytes
BYTE
FORMAT
0
Enum
Command-Specific Response Codes
CODE
CLASS
0
Success
2
3
Error
4
5
Error
6
7
Error
8-15
16
Error
17-31
32
Error
33-127
D-28
Alert Options
ON
Alert relay de-energized
Alert relay de-energized
3.75 mA
3.75 mA
DESCRIPTION
Alert Option Mode (0 - disabled, 1 - enabled)
DESCRIPTION
Alert Option Mode (0 - disabled, 1 - enabled)
DESCRIPTION
No Command-Specific Errors
Undefined
Parameter too large
Undefined
Too few data bytes
Undefined
In write protect mode
Undefined
Access Restricted
Undefined
Busy
Undefined
OFF
Alert relay energized
Alert relay energized
21 mA
21 mA

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ultima xeUltima xirUltima xa

Table of Contents