Command #185: Write Alarm Acknowledge

This command writes an Alarm Acknowledge command to the Ultima X
Gas Monitor. The alarm acknowledge command instructs the device to
clear any latched alarms in the device, provided the setpoint level for
the alarm has receded. Valid command number is on 1.
Request Data Bytes
BYTE
FORMAT
0
Unsigned
Response Data Bytes
BYTE
FORMAT
0
Float
Command-Specific Response Codes
CODE
CLASS
0
Success
1-2
3
Error
4
Error
5
Error
6-15
16
Error
17-31
32
Error
33-127
D-26
DESCRIPTION
Alarm Acknowledge command number
DESCRIPTION
Alarm Acknowledge command number
DESCRIPTION
No Command-Specific Errors
Undefined
Parameter too large
Parameter too small
Too few data bytes
Undefined
Access Restricted
Undefined
Busy