Command #186: Write Protect Mode

This command sends a single, unsigned byte to the device. Sending a
one puts the device in write protect mode. In write protect mode, all
writes and commands are ignored except a command to disable the
write protect. Only reads to the device can be made. Sending a disable,
releases the device from write protect mode. During write protect mode,
all local (Pushbutton) controls are locked out as well.
Request Data Bytes
BYTE
FORMAT
0
Enum
Response Data Bytes
BYTE
FORMAT
0
Enum
Command-Specific Response Codes
CODE
CLASS
0
Success
2
Error
3-4
5
Error
6
7
Error
8-15
16
Error
17-31
32
Error
33-127
DESCRIPTION
Write protect Mode (0 = disable, 1 = Enable)
DESCRIPTION
Write protect Mode (0 = disable, 1 = Enable)
DESCRIPTION
No Command-Specific Errors
Invalid selection
Undefined
Too few data bytes
Undefined
In write protect mode
Undefined
Access Restricted
Undefined
Busy
Undefined