Request Data Bytes
None.
Response Data Bytes
BYTE
FORMAT
DESCRIPTION
0-3
ASCII
Sensor gas type description
(see TABLE D-8)
Command #130: Read Device Real Time Clock
Reads the Real Time clock hours and minutes from the Ultima X Gas
Monitor.
Request Data Bytes
None.
Response Data Bytes
BYTE
FORMAT
DESCRIPTION
0
Unsigned
RTC Hours
1
Unsigned
RTC Minutes
Command #131: Read Alarm Setpoints
Reads the Ultima X Alarm Setpoint values.
Request Data Bytes
None.
Response Data Bytes
BYTE
FORMAT
DESCRIPTION
0-3
Float
Alarm 1 Setpoint Value
4-7
Float
Alarm 2 Setpoint Value
8-11
Float
Alarm 3 Setpoint Value
D-8