Fault Codes And Corrective Actions; Sensor Diagnostics - Siemens SITRANS FC330 Operating Instructions Manual

Coriolis flowmeters
Table of Contents

Advertisement

Info icons
Icon
Name
Read only
Data exchange
Custody transfer enabled
Memory card
Memory card
Memory card
Memory card
10.2

Fault codes and corrective actions

10.2.1

Sensor diagnostics

ID
36
37
38
SITRANS FC330
Operating Instructions, 01/2019, A5E44030648-AB
Description
● Write access disabled
● Indicate read only parameters
● Device is communicating
N/A
Capacity is 25 % used
Capacity is 50 % used
Capacity is 75 % used
Capacity is 100 % used
Icons
Message
Sensor supply volt. out of
range
Sensor supply volt. out of
range
Temperature measure‐
ment fault
Diagnostics and troubleshooting
10.2 Fault codes and corrective actions
Cause/Action
Contact your local Siemens representative.
Contact your local Siemens representative.
Turn off the power, wait 5 seconds and turn on the power
again. If the problem persists, contact your local Siemens
representative.