Parameter assignment
8.2 Functions
Note
Increased reaction time
The reaction time of the sensor increases when the process noise is damped.
8.2.6

Inputs and outputs

The hardware functionality of input and output is fixed when ordering the product. The available
configuration is described in the following table:
Channel
1
2
3
134
HW configuration
(fixed when ordering)
Current output (4-20 mA)
HART, Modbus, PROFIBUS
DP, PROFIBUS PA
Signal output
Signal output
Relay output
Signal input
SW configuration
available to the user
● Current (0/4-20 mA)
● Frequency or pulse
● Three-stage analog valve dosing control
● Discrete one or two-valve dosing control
● Operational and alarm status
● Current (0/4-20 mA)
● Frequency or pulse
● Redundant frequency or pulse (together with channel 2)
● Three-stage analog valve dosing control
● Discrete one or two-valve dosing control
● Operational and alarm status
● Discrete one or two-valve dosing control
● Operational and alarm status
● Dosing control
● Totalizer reset
● Remote zero adjust
● Force or freeze output(s)
Operating Instructions, 01/2019, A5E44030648-AB