Message
CH1 cable break
Internal error
Mass flow simulated
Volume flow simulated
Density simulated
Fraction simulated
Standard density simula‐
ted
Standard volume flow si‐
mulated
Totalizer 1 simulated
Diagnostics and troubleshooting
10.2 Fault codes and corrective actions
Cause/Action
Check current output cable connection. Passive operation:
check external power supply.
Internal error in transmitter. Turn off the power, wait 5 sec‐
onds and turn on the power again. If the error still exists,
contact your local Siemens representative.
Disable the simulation to return to normal operation.
Disable the simulation to return to normal operation.
Disable the simulation to return to normal operation.
Disable the simulation to return to normal operation.
Disable the simulation to return to normal operation.
Disable the simulation to return to normal operation.
Disable the simulation to return to normal operation.