Connecting The Phoroptor Vrx System - Ametek Reichert Phoroptor VRx User Manual

Digital refraction system
Hide thumbs Also See for Reichert Phoroptor VRx:
Table of Contents

Advertisement

Instrument Setup

Connecting the Phoroptor VRx System

WARNING: APPLY RATED INPUT VOLTAGE TO THE UNIT AS INDICATED ON THE DATA PLATE TO PREVENT
DAMAGE TO THE INSTRUMENT AND/OR INJURY TO THE OPERATOR OR PATIENT.
WARNING: CAREFULLY ARRANGE THE CABLES FOR THE UNIT AND ACCESSORIES, SO THE CABLES DO NOT
PRESENT A TRIPPING HAZARD TO THE EXAMINER OR A DANGER TO THE PATIENT.
WARNING: POSITION THIS INSTRUMENT SO THE PLUG IS EASILY ACCESSIBLE.
WARNING: DO NOT PLUG IN THE CENTRAL UNIT UNTIL THE PHOROPTOR VRX SYSTEM IS FULLY SETUP.
The basic system setup contains three components programmed to interact with each other: the
Phoroptor Head, the Central Unit, and the Controller. The Central Unit directs communication between
the Phoroptor Head, Controller, and other pre-test devices, while the Controller acts as the user interface
and controls the Phoroptor Head and the Input/Output of data.
The Phoroptor VRx is also designed to communicate with Reichert and other manufacturers acuity systems.
External devices, including digital acuity systems, projectors, Lensmeters, Auto Refractors, printers, and
computers are connected using serial cables or Bluetooth wireless serial adapters.
Note: Ensure that the pre-test equipment being interfaced is compatible. Refer to Appendix A for a
complete list of external devices that communicate with the Phoroptor VRx.
Please refer to the Connection Diagram below for plug and cable layout.
Power
16241-101 Rev. E
(continued)
Phoroptor Head
Central Unit
Figure IS-02, Connection Diagram
Projector or Digital Acuity System
EMR
Controller
11