Binocular Balance - Ametek Reichert Phoroptor VRx User Manual

Digital refraction system
Hide thumbs Also See for Reichert Phoroptor VRx:
Table of Contents

Advertisement

Performing an Exam

Binocular Balance

1. Press the BB button, or tap BB on the main screen, to begin Binocular Balance.
2. The Binocular Balance test automatically places prism in each eye (3D base down in the right eye
and 3D base up in the left eye), and fogs the left and right eye.
Note: The sphere values automatically fog by the amount specified in the Settings menu.
3. Verify that the patient can see two images, one up and one down.
Tap Prism to adjust the Prism values if the patient cannot see two images.
Tap Set Prisms to save the modified Prism values.
4. Ask the patient: "Which image is clearer, the top or the bottom?"
If the top image is clearer, increase sphere in the right eye.
If the bottom image is clearer, increase sphere in the left eye.
5. Adjust sphere until both lines look equal.
6. Press the CONTROL KNOB, or tap Equal.
7. Decrease sphere in both eyes to reach the best visual acuity.
8. Press the CONTROL KNOB, or tap Save to complete the test.
54
(continued)
16241-101 Rev. E

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents