Operation
5.4.2

With ProLink II

The Process Variables window opens automatically when you first connect to the transmitter. This
window displays current values for the standard process variables (mass, volume, density,
temperature, external pressure, and external temperature). If you have closed the Process Variables
window, select
ProLink > Process Variables
To view petroleum measurement process variables (if the petroleum measurement application is
enabled), select
ProLink > API Process Variables
To view concentration measurement process variables (if the concentration measurement application
is enabled), select
variables that are displayed depend on the configuration of the concentration measurement
application.
5.4.3

With PROFIBUS EDD

Select
View > Process Variables
concentration measurement variables are not displayed on this screen.
Select
Device > API
to view concentration measurement variables.
Variables
5.4.4

With bus parameters

To view standard process variables, examine index 26 (AI Out) of the appropriate AI function block.
Refer to Section 2.5 for information about how slots correspond to AI function blocks.
5.5

Using sensor simulation mode

Sensor simulation mode causes simulated values to be substituted for actual process data from the
sensor. Sensor simulation mode can be enabled only with ProLink II (Figure 5-1).
Configuration and Use Manual
ProLink > CM Process Variables
to view standard process variables. petroleum measurement and
to view petroleum measurement variables. Select
.
.
. The concentration measurement process
Device > CM Process
85