Emerson Micro Motion 2700 Configuration And Use Manual page 80

With profibus-pa
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

Configuration
Figure 4-14
Configuring cutoffs
Mass Flow
Mass Flow Low Cutoff
Flow tab
Enter values in the Mass Flow
Cutoff or Volume Flow Cutoff
boxes
Apply
Cutoffs
72
EDD
MMI Coriolis Flow >
Transducer Block >
Measurement >
Process Variable
Volume Flow
Volume Flow Low
(1)
Cutoff
ProLink II
ProLink >
Configuration
Enter a value in the Density
(2)
Bus parameters
Block: Transducer Block 1 (Slot 11)
Index 9 (Mass flow cutoff)
Index 39 (Liquid volume flow cutoff)
Index 40 (Density cutoff)
Index 69 (Gas standard volume flow cutoff)
Density
Density Cutoff
Density tab
Cutoff box
Apply
Model 2700 Transmitter with PROFIBUS-PA
Notes:
(1) When Gas Standard Volume is
configured, this option will be
displayed as Gas Std Vol Flow
Cutoff.
(2) When Gas Standard Volume is
configured, this box will be labeled
Std gas vol flow cutoff.