Calibration
Example
To adjust meter factors:
With the EDD, bus parameters, or ProLink II, refer to the menu flowcharts in Figure 3-8.
With the display, refer to the menu flowchart in Figure B-12.
Figure 3-8
Adjusting meter factors
ProLink II
36
The flowmeter is installed and proved for the first time. The flowmeter
mass measurement is 250.27 lb; the reference device measurement is
250 lb. A mass flow meter factor is determined as follows:
MeterFactor
The first mass flow meter factor is 0.9989.
One year later, the flowmeter is proved again. The flowmeter mass
measurement is 250.07 lb; the reference device measurement is
250.25 lb. A new mass flow meter factor is determined as follows:
MeterFactor
The new mass flow meter factor is 0.9996.
EDD
MMI Coriolis Flow >
Transducer Block
Measurement >
Process Variable
Mass Flow
Mass Factor
Bus parameters
Meter Factors
250
×
----------------- -
=
1
MassFlow
250.27
250.25
×
----------------- -
=
0.9989
MassFlow
250.07
Density
Density Factor
Block: Transducer Block 1 (Slot 11)
Index 36 (Mass factor)
Index 37 (Density factor)
Index 38 (Volume factor)
Model 2700 Transmitter with PROFIBUS-PA
=
0.9989
=
0.9996
Volume Flow
Volume Factor