Emerson Micro Motion 2700 Configuration And Use Manual page 154

With profibus-pa
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

Slave Diagnostic Response Bytes
Table E-14
Byte 14
Bit
48
49
50
51
52
53
54
55
Table E-15
Byte 15
Bit
56
57
58
59
60
61
62
63
Table E-16
Byte 16
Bit
64
65
66
67
68
69
70
71
146
Indication
Reserved (always set to 0)
Reserved (always set to 0)
Reserved (always set to 0)
Reserved (always set to 0)
Reserved (always set to 0)
Reserved (always set to 0)
Reserved (always set to 0)
Extension available
Indication
Undefined (A000)
EEPROM checksum error (A001)
RAM test error (A002)
Sensor not responding (no tube interrupt) (A003)
Temperature sensor out-of-range (A004)
Input over-range (A005)
Transmitter not characterized (A006)
Reserved
Indication
Density outside limits (A008)
Transmitter initializing/warming up (A009)
Calibration failure (A010)
Excess calibration correction, zero too low (A011)
Excess calibration correction, zero too high (A012)
Process too noisy to perform auto zero (A013)
Transmitter failed (A014)
Reserved