2. Program the separation method presented in the table on the following
page.
•
From the left side of the screen, click on the Fraction Collection
button. In the pop-up window that appears, choose Collect All
with a fraction size of 2.00 ml and a delay of 0.
•
Program the remaining steps by selecting the appropriate Add
Step icons from the left side of the screen.
BioLogicHR - Dr. J. Smith - Starter Kit - UNO Q1 -
Edit
View
Utilities
File
New
Edit
New
Browser
Method
Method
Run
Add Step
Volumn
Isocratic
Flow
1
0.0
Load/Inject
2
0.0
Sample
Linear
3
0.0
Gradient
Change
4
1.0
Valve
5
1.0
Hold
6
1.5
Alarm
2.3
7
Zero
Baseline
15.3
8
UV Lamp
18.1
9
Chart
Recorder
26.1
Repeat
Steps
Fraction
Collection
Gradient Pump
0 %B
1 psi
0.00ml/min
Enter volume of sample to load
Fig. 9. Protocol Editor for a Manual BioLogic HR System.
Options
Window
Help
1
wash
2
load
Report
Manual Setup Protocol
Description
Collect Fractions of size 2.00 ml during entire run
Chart Recorder
Isocratic Flow
Set Zero Baseline
Load/Inject Sample
Isocratic Flow
Linear Gradient
Isocratic Flow
Isocratic Flow
End of Protocol
UV
0.00375 AU
27
Run
Notes
PostRun
Log
Parameters
Turn ON
100%
25mM tris
25mM tris, 0.5M NaCl,
0%
UV Detector
Load: Sample
Static Loop
Manual Inject Valve
25mM tris
100%
0%
25mM tris, 0.5M NaCl,
25mM tris
100% --> 50%
25mM tris, 0.5M NaCl,
0% --> 50%
0%
25mM tris
25mM tris, 0.5M NaCl,
100%
25mM tris
100%
25mM tris, 0.5M NaCl,
0%
Conductivity
3.59 mS/cm
Cut
Copy
Paste
Delete
Volume: 1.0 ml
Flow: 4.00 ml/min
Volume: 0.5 ml
Flow: 4.00 ml/min
Volume: 0.8 ml
Flow: 4.00 ml/min
Volume: 13.0 ml
Flow: 4.00 ml/min
Volume: 2.8 ml
Flow: 4.00 ml/min
Volume: 8.0 ml
Flow: 4.00 ml/min
SIM1/SIG
SIM1/pH
-0.003 Volt