Figure 36 - Sub-Menu Datalogging Flow Menu; Diagnostic - Test Security - Firmware Upgrade - Ametek OXYvisor Installation, Operation & Maintenance Manual

Optical oxygen analyzer for use with bosx optical sensors
Table of Contents

Advertisement

Diag-Test
Datalogging
Security

5.14.7 - Diagnostic - Test Security - Firmware Upgrade

Firmware Upgrade
If necessary the firmware can be updated via a USB stick and initiated via this screen
5.14.8 - Diagnostic - Test Security - Reset to Factory Defaults
Reset to Factory Defaults
The analyzer can be reset to all of its original default factory settings.
5.15 Sub-Menu - Datalogging - Flow Menu
5.15.1 - Diagnostic - Datalogging
Datalogging
This sub-menu allows for the user to view the on-board error and calibration logs along with setting up and
starting the data-logger
Error Log - Shows the last 10 errors with date and time stamps.
Calibration Log - Shows the last 10 calibrations with date and time stamps.
Data Log - Start or stop the current data logging set.
Download - Either via .csv or .txt file
Data Trend Setup - Select the parameters to download (O
Deadband - Assign the the set-point as high or low. Enter the Deadband in % O2.
64
Installation, Operation and Maintenance Manual
Datalogging
Error Log
Last 10 errors
Calibration
Log
Data Log
Data Trend
Parameter to
Setup

Figure 36 - Sub-Menu Datalogging Flow Menu

OXYvisor Optical Oxygen Analyzer
Status
Start data logging
Stop data logging
Download
.csv File
Log File
.txt Delimited File
Oxygen
log
Process Temperature
Process Pressure
Phase Angle
Signal Intensity
, Temp, Press, Phase Angle, Signal Intensity).
2