Bos Sensor With External Rtd's Via J-Box; Bos Sensor With External Rtd Via Rear (Wiring) Compartment; Figure 16 - Bos Sensor And External Rtd Through Junction Box; Figure 17 - Bos Sensor And External Rtd Or Zone 1 - Ametek OXYvisor Installation, Operation & Maintenance Manual

Optical oxygen analyzer for use with bosx optical sensors
Table of Contents

Advertisement

Installation, Operation & Maintenance Manual
OXYvisor Optical Oxygen Analyzer

4.10.2 BOS Sensor with External RTD's via J-Box

The BOS sensor is commonly specified without an integral RTD for gas phase applications. An external RTD
can be placed directly into the process and is preferred for accurate temperature measurement under various
process conditions (e.g. process with fast moving temperature, low heat capacity (dry gas) or TProcess -
TAmbient is large). The external RTD can be connected through the J-Box (Fig 16) or through the Rear
(Wiring) Terminal.

4.10.3 BOS Sensor with External RTD via Rear (Wiring) Compartment

For OXYvisor installations in Zone 1, where the model is selected with Agency Approval "1", the BOS
sensor (as standard) is not allowed to be installed with integral RTD. An external RTD will be the standard
recommended installation. For this case, the RTD can be wired through the Rear (Wiring) Compartment as
shown in Figure 17.
Oxygen
01/06/2017 10:55 AM
% O2
850.9 mBar
20.946
67.9°F
Err: 0
OXYvisor
ESC
ANALYZER FIBER
OPTIC SMA
CONNECTION IN
J-BOX
FIBER OPTIC SMA
TO SMA COUPLER
P/N: B3907-1122
BOS OXYGEN
SENSOR FIBER
OPTIC SMA
CONNECTION IN
J-BOX

Figure 16 - BOS Sensor And External RTD through Junction Box

Oxygen
01/06/2017 10:55 AM
% O2
850.9 mBar
20.946
67.9°F
Err: 0
OXYvisor
ESC
ANALYZER FIBER
OPTIC SMA
CONNECTION IN
J-BOX
FIBER OPTIC SMA
TO SMA COUPLER
P/N: B3907-1122
BOS OXYGEN
SENSOR FIBER
OPTIC SMA
CONNECTION IN
J-BOX

Figure 17 - BOS Sensor And External RTD or Zone 1

EXAMPLE - GENERAL ARRANGEMENT
OF OXYVISOR (General Purpose / Class I
Div. 2 / or Zone II 3G) and BOS SENSOR
WITH EXTERNAL RTD
ANALYZER RTD
MOLEX CONNECTION
IN J-BOX
BOS OXYGEN SENSOR
RTD MOLEX CONNECTION
IN J-BOX
BOS OXYGEN SENSOR
WITHOUT RTD
(SEE BOS DATA
SHEET FOR OPTIONS)
EXAMPLE - GENERAL ARRANGEMENT
OF OXYVISOR (Zone I 2G) and BOS SENSOR
WITH EXTERNAL RTD
RIGID CONDUIT or Ex d GLANDS & CABLING
USER SUPPLIED RTD OR
TEMPERATURE TRANSMITTER
CERTIFIED FOR HAZARDOUS
AREA
BOS OXYGEN SENSOR
(SEE BOS DATASHEET
FOR OPTIONS)
EXTERNAL RTD
PT1000 RTD
BOS-FLEXSENSE
BOS-FLEXSENSE
Section 4
Installation
33

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents