A7.5 - Communication Examples

A7.5.1 Read Process Variables:
To interrogate process variables such as oxygen, phase angle and amplitude
80
Installation, Operation and Maintenance Manual
First Register
Function
Code
(Minus Offset = 40000)
40130
6
40131
6
40132 (1st Schedule)
40142 (2nd Schedule)
40152 (3rd Schedule)
6
40162 (4th Schedule)
40138(1st Schedule)
40148(2nd Schedule)
40158(3rd Schedule)
16
40168(4th Schedule)
40140(1st Schedule)
40150(2nd Schedule)
40160(3rd Schedule)
16
40170(4th Schedule)
40175
6
40172 (Pressure Value)
16
40174 (Pressure Unit)
6
40177 (Oxygen Value)
16
40176 (Oxygen Unit)
6
40179 (Oxygen Value)
16
40181 (Oxygen Unit)
6
40182 (Oxygen Value)
16
40184 (Oxygen Unit)
6
40185
16
40187
16
40189
6
44000
6
OXYvisor Optical Oxygen Analyzer
Data Type
Uint16
1=Dry Gas; 2=Humid
Uint16
1=Enable; 2=Disable
Uint16
1=Enable; 2=Disable
Enter: 0 – 5184000
Uint32
(0-60 days in seconds)
Enter: Current UTC time in seconds + start time in seconds
Uint32
within next 60 days. Refer to Step #5 above- Page
Uint16
1=Ambient Sensor; 2=Manual Pressure Input
1=mbar; 2=inch H20; 3=bar; 4=Torr;
Float
Uint16
5=PSI
Float
1=% Oxygen; 2=PPM
Uint16
Float
1=% Oxygen; 2=PPM
Uint16
Float
Uint16
1=% Oxygen; 2=PPM
Float
Enter: 1-30(in mins)
Float
Enter: 1-30(in mins)
Uint16
Enter: 1-120(in secs)
Uint16
4=Start ; 5=Stop
Value to Write
XYZ