A7.5.3 Triggering Test Gas Insert (TGI) - Remote Validation

A7.5.4 Setting Up Test Gas Insert
82
Installation, Operation and Maintenance Manual
If the TGI has already
been previously setup
and you want to simply
trigger it, write 1 to
register 44009
First set up the "Time
period". If you want
to TGI to run for 2
minutes, write 2 to
register 40408.
OXYvisor Optical Oxygen Analyzer
If the TGI has not been configured, by default the
"Time period" = 3 mins and the "Output Gas" is set
to Zero
Gas. Also, the TGI will not trigger if the
Auto-Calibration has not been enabled.
USEFUL INFORMATION
The test gas will
automatically stop
after the "time
period" set during the
configuration of the
TGI. If you want to
stop the TGI before the
time period is up; write
value 2 to the same
register 44009.
To define the Output
gas as "Span Gas",
write 2 to register
40410.
Once the TGI has
been configured to
the desired time and
output, follow steps in
section 5.3 to trigger it.