Oxyvisor Configuration - Product Model Selection; 2�4 Oxyvisor Accessories - Ametek OXYvisor Installation, Operation & Maintenance Manual

Optical oxygen analyzer for use with bosx optical sensors
Table of Contents

Advertisement

Section 2
Technical Product Specification

2.3 OXYvisor Configuration - Product Model Selection

Agency
Analyzer
Power
Approval
OXYvisor Base Model Number
BOA
OXYvisor - Barben Oxygen Analyzer
Input Power
DC
22 to 26.5 VDC, 5W (4-wire, line powered analyzer, this is NOT a loop powered analyzer. Requires two wires
for DC power and two separate wires for 4-20 mA output)
AC
85 to 264 VAC, 47-63 Hz, 6 W (4-wire, line powered analyzer)
Agency Approval
Agency
Analyzer
Power
Approval
BOA
DC
Table 3 - OXYvisor Configuration - Product Model Selection
2�4 OXYvisor Accessories
Part Number
B5008-112x
B5008-1125
B5012-
B4000-1
B5500-1
B5500-2
BOS Data Sheet
(link)
SCP Data Sheet
(link)
Table 4 - OXYvisor Accessories Parts
14
Installation, Operation and Maintenance Manual
Sensor
Mounting
Style
Orientation
1
II 2 G
Class I Zone 1
2
Class I Division 2
II 3 G
Sensor Style
SFP
Standard Fiber Patch
-
Saved for future use - Integral Wands with lengths (2.5, 5.0 and 10 M)
Mounting Orientation
B
-
Sensor
Mount
Style
Orientation
2
SFP
B
Wall Mount Kit - 316 SS with Swivel
Pipe Mount Kit - 316 SS (1-1/2" - 2-1/4" pipe)
Compact USB memory stick for data logging and firmware upgrades, 8 GB
Compact sunshade, powder coated, highly recommended for outdoor installations
Trace Level - AutoCal Kit: Test Gas Bottles (N6 & 25 ppm), analytical grade regulators, pipe/wall
stand for bottles, (requires AutoCal SCP) [customer supplied 1/4" 316 SS tubing required]
Trace Level - AutoCal Kit: Test Gas Bottles (N6 & 50 ppm), analytical grade regulators, pipe/wall
stand for bottles, (requires AutoCal SCP) [customer supplied 1/4" 316 SS tubing required]
For BOS Sensor Selection see BOS Specification Data Sheet
For Sample Calibration Panel options, AutoCal or Manual, see SCP Specification Data Sheet
OXYvisor Optical Oxygen Analyzer
Conduit
Entries
Ex db op is IIC T4 Gb
Ex db op is IIC T4 Gb
AEx db op is IIC T4 Gb
Ex db op is IIC T4 Gb
Group A, B, C, D T4a
Ex ec ic op is IIC T4 Gc
Junction Box placed below main enclosure, fiber optic exits bottom
Saved for future use - other orientations w/ display and JB
Conduit Entries
SI
25 mm Conduit Entries
AM
3/4" FNPT Conduit Entries
Conduit
Entries
AM
Description
//
ATEX
//
IEC / EU
//
US (NEC 505)
//
CA (CEC Section 18)
//
US (NEC 500) and CA (CEC Annex J18)
//
PENDING CERTIFICATION FROM NRTL
Typical Analyzer Model Number
(as shown)

Hide quick links:

Advertisement

Table of Contents
loading